Sealing: To seal the donor tube, a dielectric or RF sealer such as the
SEBRA model 2600/1105 Hand Held sealer is recommended, Click
here for a link to the SEBRA website. The sealer should be operated
and maintained according to manufacturer's instructions to ensure
optimal performance. |
![]() |
![]() |
Controlled rate freezing to below -40ºC is highly recommended to help prevent breaking of the bag in storage. The use of a fixed rate freezing protocol is strongly suggested to freeze the bags before placing into the vapor phase of a liquid nitrogen storage tank. |
1. Prepare: Dry the freezing bag and the interior of a freezing
cassette. It is recommended that only one bag be stored per cassette.
2. Freeze: Freeze the bag in the cassette using a controlled rate
freezing protocol. Freezing the bag to below -40ºC.
DMSO Cryoprotectant Procedure* Click
here for link
DMSO/Pentastarch Cryoprotectant Procedure* Click
here for link
*The above procedures were presented at the ISCT 2001 meeting; these sample procedures are for reference only.
Following completion of the freezing program, remove the cassettes from
the freezing chamber and place them into the storage racks within the
vapor phase of the nitrogen storage tank. Transfer the bags to the storage
tank as quickly as possible.
A uniform thickness of the product is recommended for best freezing results.
The fill volume is dependent on the way that the bag is to be stored in
liquid nitrogen. This slight variance is best determined by the user.
Cryostore Freezing Bags are fragile when frozen, please handle carefully.
While the Cryostore Freezing bags are suitable for direct immersion in
the liquid phase of the nitrogen tank, it is recommended that the container
be stored in the vapor phase of the tank to avoid possible cross contamination.
Never move room temperature products into Liquid Nitrogen, as bursting
of the bag is very probable. Do not immerse the CryoStore freezing bag
directly into liquid nitrogen from the temp control freezer. Allow to
cool to vapor phase temperatures first. Anything that causes extreme thermal
shock can result in damage to the Cryostore bag.
Always follow your institutional protocol and all safety procedures when
working with liquid nitrogen.
The Cryostore bag has been designed to fit into existing, standard cassettes.
Check to be sure that the selected bag fits comfortably in the selected
cassette. We recommend using one bag per cassette.
| OriGen Biomedical 4020 S. Industrial Dr. Suite 160 Austin, Texas USA 78744 Tel: +1 (512) 474-7278 Fax: +1 (512) 617-1503 Copyright ©: 2002-2007, OriGen Biomedical |
OriGen Biomedical Europe |
Key list: Blood banking, cord blood storage, cryopreservation,stem cell storage, cord blood freezing, blood component freezing, hematopoetic